Minimum Order Quantity | 1 Unit |
Application | Clinical Purpose, Hospital |
This lightweight PLV-102 is well suited for transport and portable applications. Controlled by a microprocessor, the piston-driven PLV-102 delivers volume ventilation in a range of settings. With features that include three power source options, it is ideal for use in the sub-acute care, nursing home, or home settings.
Front Panel Controls:
- Mode Selection: Control, Control + Sigh, Assist/Control, Assist/Control + Sigh, SIMV
- Tidal Volume: 0.05 to 0.20 ± 0.02 liters; 0.20 to 3.00 liters ± 10%
- Sign Volume: 150% of Set Tidal Volume
- Sign Frequency: 1 Sign Breath every 100 Tidal
- Breath Rate: 2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2
- Inspiratory Flow Rate (Peak flow rate during inspiratory time): 10 to 120 LPM
- Battery Test Switch: Test Internal/External Battery
- Main Power Switch: On/Off, Recharge
- DC Circuit Breaker: Push/Pull Actuation
- AC Circuit Breaker: Push/Pull Actuation
- Airway Pressure Limit: 5 to 100 cm H2O ± 5
- Sensitivity: +18 cm H2O ± 1 to at least -6 cm H2O
- Low Pressure Alarm: 2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5
- Manual Sigh: Push to Activate One Breath
- Oxygen Percent (See Operators Manual for Oxygen Delivery Limitations): 21% to 90% ± 6%
- Oxygen Sensor – SIMV Operation (See Operators Manual for Oxygen Delivery Limitations): 21% to 90% ± 10%
Power Sources:
- 120 VAC 50/60 Hz
- 12 VDC External Battery
- 12 VDC Internal Battery
Physical Dimensions:
- Overall Dimensions: 9″ H x 12.25″ W x 12.25″ D (22.9 cm x 31.1 cm x 31.1 cm)
- Weight: 28.9 lb (13.1 kg)