It’s an advanced analyzer for Complete Hemostasis. This four channel coagulation analyzer redefines the benefits in its class. Erba ECL 412, can perform PT, APTT, Fibrinogen, Thrombin Time , All Factors, Lupus , Protein S, Protein C , Antithrombin III and DDimer.
Product Features:
- Clotting Tests such as PT, aPTT, Fibrinogen, Thrombin time, Factors, Lupus anticoagulant, Protein S performed using Light scatter principle.
- Protein C, Antithrombin III done by Chromogenic Principle
- Tests like D Dimer performed using Immuno-turbidimetry principle
- Interactive touch screen interface
- Clotting Graph provides details of the clotting test
- Cuvette and reagent management software
- Twelve levels of QC with Levy-Jennings Plot- Identifies random and systematic errors
- Temperature stability at 37 degree C to perform the tests
- Low Reagent Consumption
- USB Connectivity
- CE mark
Product Applications:
- ECL 412 has uses in various routine IVD labs for doing routine PT and aPTT
- It has application in Blood Banks for Quality Control of Fibrinogen and Factor VIII
- It is used as Standby to bigger systems, in accredited centers
- It is used in Research Facilities to perform PT and aPTT on animal samples such as rat, mouse, rabbit, hamsters etc
Technical Specification
General Specifications:
- 4 channel semi-automated hemostasis instrument
- It can perform 4 simultaneous tests of the same parameter
- Sample id can be 14 digits alphanumeric, with progressive id or custom id options
- It is also possible to have a second row of assay preparation/incubation during the analysis of the current line, using preparation line option
- It can run assays in single or duplicate (programmable per assay)
- Optically detects and automatically moves to the next assay steps and auto starts the measurements
Tests principles:
- Clotting (scattered light at 640nm)
- Chromogenic (colorimetry at 405nm)
- Immuno-turbidimetric (turbidimetry at 800nm)
- Dimensions / Weight:
- Dimension : 300 x 290 x 90 mm
- Weight : 3 kgs
Environmental requirements:
- Operating Temperature : 17-32°C
- Humidity: Max. 80% Relative Humidity, non-condensing
Power:
- Power Supply: 100-240 VAC and 50/60 Hz
- Power Consumption: 45 Watts
- In-Rush Power: 150VA or less
Calibrations:
- Up to 6 calibration levels per assay
- Storage of the active calibration curve
Quality Controls:
- Up to 12 different control names total can be assigned to any given assay
- Automatic Levey-Jennings plotting
- Automatic statistics calculation
- Storage of Quality Control results
Stored results:
- Independent storage allocation in the instrument memory for Quality Control (QC) and patient determinations
- Up to 600 results for QC determinations (run in duplicate)
- Up to 900 results for patients (run in duplicate)
- New results automatically replace the oldest results of the appropriate category past the maximum storage allowance
- Reaction curves are exclusively stored on the external USB pendrive.
- Regular backup and file purging is recommended from the USB pendrive to insure rapid operations. (We recommend at least a monthly purge from a computer to back up and remove CH files)
- USB pendrive , FAT32 formatted