Product Information
Daytona Plus Provides Eyecare Professionals With UWF Digital Images Of 200 Degrees Or Up To 82% Of The Retina In A Single, Non-contact Optomap Image. In Addition, The Daytona Plusdevice Comes With The OptosAdvance™ Browser-based Image Review Software, Which Allows For Simple Documentation, Monitoring, And Referral Processing To Assist In Patient Management And Improved Patient Flow.
Product Description
Why Daytona plus?
Daytona plus provides eyecare professionals with UWF digital images of 200 degrees or up to 82% of the retina in a single, non-contact optomap image. In addition, the Daytona plus device comes with the OptosAdvance™ browser-based image review software, which allows for simple documentation, monitoring, and referral processing to assist in patient management and improved patient flow.
Daytona plus imaging modalities and image viewing options:
Image Modalities
– Color
– Red-free
– Choroidal
– Autofluorescence
Image Views
– Standard: 200⁰ Single Capture
– Auto-montage: Up to 220⁰
– Central Pole: Detailed view of the macula
– Stereo: Image pairing for optic disc and retinal evaluation
Technical Specifications
Image Modalities:
optomap color and optomap plus (red and green laser):
Color composite view
Green laser view
Red laser view
optomap af (green laser):
Autofluorescence
Resolution
optomap color: 20 μm
optomap plus, af : 14 μm
Wavelengths
Red laser: 635 nm
Green laser: 532 nm
Exposure Time
Less than 0.4 seconds
Footprint
Width: 440 mm/18 in
Depth: 500 mm/20 in
Height: 795 mm/32 in
Weight
28 kg/62 lbs
Table Space Requirements
Width: 700 mm/28 in
Depth: 550 mm/22 in
Colors
Black body with red trim
White body with red trim
White body with silver trim
Laser Class
Laser safety class-1
following EN60825
System Voltage
US: 100-120V at 50/60Hz 3a
Power Consumption
Max. 300VA
color
Red Free
Choroidal
Autofluorescence