Product Details:
Display: Digital
Model: M747
Automation Grade: Automatic
Size: 12.1 TFT
I Deal In: New Only
Brand: Meditech England
Specification
Features:
The choice of 12.1” TFT color display provides superior flexibility for different clinical applications
Accurate monitoring of ECG, SPO2, NIBP, TEMP, RESP, HR, PR and optional IBP, EtCo2
Pacemaker detection, Electrosurgical interference proof
Defibrillation protection and defibrillation synchronization
OxyCRG available to judge the respiration and circular function for neonates
120 hours graphic and tabular trends of all parameters
USB ports and SD card slot to ensure data storage/function upgrades
Select from either Sidestream, or Mainstream EtCO2 technology to obtain fast, simple, and accurate CO2 results
Nurse call function and Bi-directional communication with Central
Display:
Various working interface selectable: Standard Monitoring Display, Large Font Intensive Care Display, Trend Graph/ Monitoring Co-Display, Bed to Bed view Display (optional), OxyCRG Dynamic View Display, Drug Dose Calculation Interface
Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG:
Lead Mode: 3 Leads (R, L, F or RA, LA, LL), 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Waveform: 7 waveforms display on a screen
4 selectable plus: 0.25, 0.5, 1, 2 cm/mV
PACE detection Detection leads: 7 selectable leads: I, II, III, AVR, AVL, AVF, V
Defibrillation energy distributary Input voltage range ±8mV, polarized voltage: ±500mV
CMRR Diagnosis: >100 dB (no 50Hz/60Hz software wave trap)
Monitor: >110 dB (has 50Hz/60Hz software wave trap)
Surgery: >100dB (50Hz/60Hz software wave trap)
System noise: <30μV p-p (RTI)
HR Measuring and Alarm Range
Adult/Ped: 15 bpm – 300bpm
Neo: 15 bpm – 350bpm
Accuracy: ±1% or ±1bpm, which great
Resolution: 1 bpm
Sensitivity: > 200 μV P-P
Differential Input Impedance: ≥ 5 MΩ
Electrode offset potential: 300mVd.c. – 600mVd.c.
Patient leakage current: < 10 uA
ECG Signal Range: 6 mV (Vp-p).
ST Segment Monitoring Range Measure and Alarm: -2.0 – +2.0 mV
Respiration:
Method: Impedance between R-F (RA-LL), R-L (RA-LA)
Respiration Rate Measuring and Alarm Range:
Adult: 0 rpm – 120 rpm Neo/Ped 0 rpm – 150 rpm
Resolution: 1 rpm Accuracy 2 rpm
Gain Selection: 0.25, 0.5, 1, 2, 3, 4, 5
IBP:
Channel: Upto 4 channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Static Pressure Measuring Range: (-)50 – +300 mmHg (up to 350 mmHg)
Static Pressure Accuracy: ±2% or ±1 mmHg whichever is higher
Dynamical Pressure Measuring Range: (-)50 – +300 mmHg
Dynamical Pressure Accuracy: 2 % or 1mmHg which bigger
ART: 0 – 300 (mmHg) PA 6 – 120 (mmHg)
CVP/RAP/LAP/ICP: (-)10 – 40 (mmHg) P1/P2 (-)50 – 300 (mmHg)
Resolution: 1 (mmHg) Zero range 200 mmHg